Nonmechanical spectral domain optical coherence tomography using an electrowetting beam-scanner Journal Article uri icon

Overview

abstract

  • In this work, we demonstrate the first use of an electrowetting beam-steering prism to perform lateral scanning in a benchtop spectral domain optical coherence tomography system. Electrowetting-based optics are compact, consume very little power, and have entirely nonmechanical actuation, making them ideal for small-scale applications like endoscopic OCT. Our electrowetting prism is capable of scanning at ±6° with an applied voltage of 67 V ±20~V. These capabilities are used to perform lateral scanning in a benchtop spectral domain optical coherence tomography system utilizing a 1091 nm, 53 nm bandwidth gain-managed nonlinear fiber amplifier source. We characterize the field of view, axial resolution, and sensitivity of our system and find that our electrowetting prism allows us to acquire OCT images without sacrificing system performance.

publication date

  • August 25, 2025

Date in CU Experts

  • August 20, 2025 8:56 AM

Full Author List

  • Gilinsky SD; Miscles EJ; Bartos J; Musgrave J; Huang S-W; Bright VM; Gopinath JT

author count

  • 7

Other Profiles

Electronic International Standard Serial Number (EISSN)

  • 1094-4087

Additional Document Info

start page

  • 35604

end page

  • 35604

volume

  • 33

issue

  • 17